Dickson Support Database
Oceaview Cobalt X Firmware Update Notes
| 1/23/24 | v2.10.2 | Bug Fix: #15374 -> Swap sensor doesn’t work with remote sensor (swap fails for Emerald /Atals devices) #15050 -> RAK CN470 AT+SLEEP command waits for an answer but does not give any, which freezes in Info menu |
| 10/16/23 | v2.10.1 | Bug Fix: 14771 Siren LoRa / Cobalt X 7mA consumption on battery with RAK CN470 14899 X2 – Screensaver stuck when displayed on PIN screen + C class |
| 8/24 TBR | V2.10 | Allow for alarm pause #13597 -> Pause Alarm/Warning from the data logger screen (in sync with web app v1.13) #13966 -> Incorrect unit displayed in stop when we use F instead of C #14189 -> Allow DFU with ongoing datalogging when in AutoStart Mode (Dickson mode) #14477 -> ACK button not displayed event though config is OK #14225 -> Set calibration drift and uncertainty from OCEAView web app on 0-10V Smart set wrong parameters when using mapping #13723 -> LoRa factory frequency cannot be set from Bluetooth #13471 -> Add support to China frequency CN470 #13751 -> RAK 1.0.3 and 1.0.4 with AU915 frequency, Class C, doesn’t communicate #13714 -> LoRa Configuration – Israel and Russia cannot be returned in RES_LORA_CONFIG (Get supported frequencies) #12964 -> Change “oceasoft.com” for “dicksondata-debug.freshy.dev” and remove the hyperlink format #13964 -> Add a “*” next to the date in information screen when date is not set #13853 -> Change the order of Informations#14121 -> Change blinking behavior of the Cobalt X#13854 -> Change the Settings order |
| 8/3/2023 | V2.9.1 | Bug Fix: Launch a Datalogging on CX Configure one warning threshold and one alarm threshold Try to read datalogging with OceaView mobile The warning threshold does not appear as a warning but like an alarm (red instead or orange) |
| 4/6/2023 | V2.9 | allows Warning to show as orange on the logger dipslay. If using an older version of firmware the warning will still show as red – NOTE: AS OF NOW ALL CX2 LOGGERS WILL SHIP WITH V2.8 NOT V2.9 – CLOUD ACCOUNTS WILL NEED TO UPGRADE TO V2.9 FREQUENCY – users cannot change the LoRaWAN frequency on Cobalt X data loggers. Frequencies are assigned by Dickson’s production team according to the destination region. – Upgrading an exising (v2.8 or lower) CX2 to V2.9 will lock the logger into the existing frequency – downgrading a logger to V2.8 will allow the user to change the frequency – We will not sell each frequency version under a different part number. Production will install firmware for the enduser required frequency at Dickson before shipping |